Purpose The purpose of this study was to determine the quantitative variability of diffusion weighted imaging and apparent diffusion coefficient values across a large fleet of MR systems. Using a NIST traceable magnetic resonance imaging diffusion phantom, imaging was reproducible and the measurements were quantitatively compared to known values.
Methods A fleet of 23 clinical MRI scanners was investigated in this study. A NIST/QIBA DWI phantom was imaged with protocols provided with the phantom. The resulting images were analyzed and ADC maps were generated. User-directed region-of-interests on each of the different vials provided ADC measurements among a wide range of known ADC values.
Results Three diffusion phantoms were used in this study and compared to one another. From the one-way analysis of the variance, the mean and standard deviation of the percent errors from each phantom were not significantly different from one another. The low ADC vials showed larger errors and variation and appear directly related to SNR. Across all the MR systems and data, the coefficient of variation was calculated and Bland-Altman analysis was performed. ADC measurements were similar to one another except for the vials with the lower ADC values, which had a higher coefficient of variation.
Conclusion ADC values among the three phantoms showed good agreement and were not significantly different from one another. The large percent errors seen primarily at the low ADC values were shown to be a consequence of the SNR dependence and very little bias was observed between magnetic strengths and manufacturers. ADC values between diffusion phantoms were not statistically significant. Future investigations will be performed to study differences in magnetic field strength, vendor, MR system models, gradients, and bore size. More data across different MR platforms would facilitate quantitative measurements for multi-platform and multi-site imaging studies. With the increasing usage of diffusion weighted imaging in the clinic, the characterization of ADC variability for MR systems provides an improved quality control over the MR systems.
Competing Interest Statement
Dr. Fuller has received direct industry grant support, honoraria, and travel funding from Elekta AB unrelated to this project. Part of the research was conducted at the Center for Academic Biomedical Imaging at the University of Texas MD Anderson Cancer Center with equipment support from GE Healthcare.
Funding Statement
Dr. Fuller received/receives funding and salary support related to this project during the period of study execution from: the National Institutes of Health (NIH) National Institute of Biomedical Imaging and Bioengineering (NIBIB) Research Education Programs for Residents and Clinical Fellows Grant (R25EB025787-01); the National Institute for Dental and Craniofacial Research Establishing Outcome Measures Award (1R01DE025248/R56DE025248) and Academic Industrial Partnership Grant (R01DE028290); NCI Early Phase Clinical Trials in Imaging and Image-Guided Interventions Program (1R01CA218148); an NIH/NCI Cancer Center Support Grant (CCSG) Pilot Research Program Award from the UT MD Anderson CCSG Radiation Oncology and Cancer Imaging Program (P30CA016672); an NIH/NCI Head and Neck Specialized Programs of Research Excellence (SPORE) Developmental Research Program Award (P50 CA097007). Dr. Fuller received/receives funding and salary support unrelated to this project during the period of study execution from: NIH Big Data to Knowledge (BD2K) Program of the National Cancer Institute (NCI) Early Stage Development of Technologies in Biomedical Computing, Informatics, and Big Data Science Award (1R01CA2148250; National Science Foundation (NSF), Division of Mathematical Sciences, Joint NIH/NSF Initiative on Quantitative Approaches to Biomedical Big Data (QuBBD) Grant (NSF 1557679); NSF Division of Civil, Mechanical, and Manufacturing Innovation (CMMI) grant (NSF 1933369); and the Sabin Family Foundation. Direct infrastructure support is provided to Dr. Fuller by the multidisciplinary Stiefel Oropharyngeal Research Fund of the University of Texas MD Anderson Cancer Center Charles and Daneen Stiefel Center for Head and Neck Cancer and the Cancer Center Support Grant (P30CA016672) and the MD Anderson Program in Image-guided Cancer Therapy. Dr. Fuller has received direct industry grant support, honoraria, and travel funding from Elekta AB unrelated to this project. Part of the research was conducted at the Center for Academic Biomedical Imaging at the University of Texas MD Anderson Cancer Center with equipment support from GE Healthcare. This research is supported in part by the National Institutes of Health through MD Anderson’s Cancer Center Support Grant CA016672.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
This IRB exempt phantom measurement study did not use human nor animal data, but rather, physical measurement of electronic system (MRI) performance.
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Raw data are made available as a .XLS file via figshare doi with a 12-month embargo due to pending manuscript(s) under review; DICOM files are to be made publicly available after manuscript acceptance.