Skip to main content
medRxiv
  • Home
  • About
  • Submit
  • ALERTS / RSS
Advanced Search

Saliva as a testing specimen with or without pooling for SARS-CoV-2 detection by multiplex RT-PCR test

View ORCID ProfileQing Sun, Jonathan Li, Hui Ren, Larry Pastor, Yulia Loginova, Roberta Madej, Kristopher Taylor, View ORCID ProfileJoseph K. Wong, Zhao Zhang, Aiguo Zhang, View ORCID ProfileChuanyi M. Lu, View ORCID ProfileMichael Y. Sha
doi: https://doi.org/10.1101/2020.10.27.20219196
Qing Sun
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Qing Sun
  • For correspondence: qsun{at}diacarta.com mark.lu{at}va.gov msha{at}diacarta.com
Jonathan Li
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Hui Ren
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Larry Pastor
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Yulia Loginova
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Roberta Madej
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Kristopher Taylor
2University of California and VA Healthcare System, San Francisco, CA 94121
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Joseph K. Wong
2University of California and VA Healthcare System, San Francisco, CA 94121
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Joseph K. Wong
Zhao Zhang
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Aiguo Zhang
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Chuanyi M. Lu
2University of California and VA Healthcare System, San Francisco, CA 94121
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Chuanyi M. Lu
  • For correspondence: qsun{at}diacarta.com mark.lu{at}va.gov msha{at}diacarta.com
Michael Y. Sha
1DiaCarta Inc. 2600 Hilltop Dr. Richmond, CA 94806
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Michael Y. Sha
  • For correspondence: qsun{at}diacarta.com mark.lu{at}va.gov msha{at}diacarta.com
  • Abstract
  • Full Text
  • Info/History
  • Metrics
  • Data/Code
  • Preview PDF
Loading

Abstract

Background Sensitive and high throughput molecular detection assays are essential during the coronavirus disease 2019 (COVID-19) pandemic, caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The vast majority of the SARS-CoV-2 molecular assays use nasopharyngeal swab (NPS) or oropharyngeal swab (OPS) specimens collected from suspected individuals. However, using NPS or OPS as specimens has apparent drawbacks, e.g. the collection procedures for NPS or OPS specimens can be uncomfortable to some people and may cause sneezing and coughing which in turn generate droplets and/or aerosol particles that are of risk to healthcare workers, requiring heavy use of personal protective equipment. There have been recent studies indicating that self-collected saliva specimens can be used for molecular detection of SARS-CoV-2 and provides more comfort and ease of use for the patient. Here we report the performance of QuantiVirus™ SARS-CoV-2 multiplex test using saliva as the testing specimens with or without pooling.

Methods Development and validation studies were conducted following FDA-EUA and molecular assay validation guidelines. Using SeraCare Accuplex SARS-CoV-2 reference panel, the limit of detection (LOD) and clinical evaluation studies were performed with the QuantiVirus™ SARS-CoV-2 multiplex test. For clinical evaluation, 85 known positive and 90 known negative clinical NPS samples were tested. Additionally, twenty paired NPS and saliva samples collected from recovering COVID-19 patients were tested and the results were further compared to that of the Abbott m2000 SARS-CoV-2 PCR assay. Results of community collected 389 saliva samples for COVID-19 screening by QuantiVirus™ SARS-CoV-2 multiplex test were also obtained and analyzed. Moreover, saliva pooling with 6 and 12 samples together were also evaluated.

Results The LOD for the QuantiVirus™ SARS-CoV-2 multiplex test was confirmed to be 100-200 copies/mL. The clinical evaluation using contrived saliva samples indicated that the positive percentage agreement (PPA) of the QuantiVirus™ SARS-CoV-2 multiplex test is 100% at 1xLOD, 1.5xLOD and 2.5xLOD. No cross-reactivity was observed for the QuantiVirus™ SARS-CoV-2 multiplex test with common respiratory pathogens. Testing of clinical samples showed a positive percentage agreement (PPA) of 100% (95% CI: 94.6% to 100%) and a negative percentage agreement (NPA) of 98.9% (95% CI: 93.1% to 99.9%). QuantiVirus ™SARS CoV-2 multiplex test had 80% concordance rate and no significant difference (p=0.13) in paired saliva and NPS specimens by Wilcoxon matched pairs signed rank test. Positive test rate was 1.79% for 389 saliva specimens collected from the communities for COVID-19 screening. Preliminary data showed that saliva sample pooling up to 6 samples for SARS-CoV-2 detection is feasible (sensitivity 94.8% and specificity 100%).

Conclusion The studies demonstrated that the QuantiVirus™ SARS-CoV-2 multiplex test has a LOD of 200 copies/mL in contrived saliva samples. The clinical performance of saliva-based testing is comparable to that of NPS-based testing. Pooling of saliva specimens for SARS-CoV-2 detection is feasible. Saliva based and high-throughput QuantiVirus™SARS-CoV-2 multiplex test offers a highly desirable test during the ongoing COVID-19 pandemic.

Introduction

A novel coronavirus, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2, previously provisionally named 2019 novel coronavirus or 2019-nCoV), has been identified as the cause of respiratory infection including severe pneumonia outbreak that started in Wuhan, China in late 20191-2, and has since become a global pandemic. The disease was named the coronavirus disease of 2019 (COVID-19) by the World Health Organization in February 2020. It has been determined that SARS-CoV-2 can be transmitted from person-to-person (symptomatic or asymptomatic) and is more transmissible than SARS-CoV 3-5.

Nasopharyngeal swab (NPS) and oropharyngeal swab (OPS) samples are widely accepted as specimens for the detection of SARS-CoV-2 since the start of the COVID-19 pandemic. However, the collection procedures for NPS and OPS specimens may cause discomfort and, in some people, sneezing and coughing. The latter in turn can generate droplets or aerosol particles that place healthcare workers collecting these specimens at risk,6 requiring heavy use of personal protective equipment (PPE). Poor tolerability of NPS and OPS sampling can result in false-negative tests due to inadequate or poor quality of specimen collection7-10. Recent investigations by Wyllie et al11 and Hanson et al12 suggested that saliva is a viable and even more sensitive alternative to NPS specimens, and could also enable at-home self-administered sample collection for large-scale SARS-CoV-2 molecular testing. Other researchers also reported that SARS-CoV-2 was detected in 91.7% (n=11) of the initial saliva specimens from confirmed COVID-19 patients. All saliva specimens (n=33) collected from patients whose NPS specimens tested negative for COVID-19 also tested negative13. It is apparent that detection of SARS CoV-2 in saliva can be used as an alternative, more appealing and cost-effective procedure for the diagnosis of COVID-19. Indeed, a molecular test using saliva samples was first approved for FDA under EUA on May 8, 2020.14

The use of saliva specimens might decrease the risk of nosocomial transmission of COVID-19 and is ideal for situations in which NPS or OPS specimen collection may be impractical.15-18 Collecting saliva is easy and more tolerable to patients, can reduce risk of cross-infection, and can be used in settings where PPE is not readily available. It will also be useful for testing infants and young children in daycare facilities and schools.

The QuantiVirus™ SARS-CoV-2 Multiplex Test is a real-time reverse transcription polymerase chain reaction (RT-qPCR) test that includes the assay controls for the qualitative detection of viral RNA from SARS-CoV-2 in NPS, OPS, saliva or sputum specimens collected from patients who are suspected of COVID-19 infection. Extracted RNA is reverse-transcribed and amplified in a single reaction. In this multiplex qPCR method, the Orf1ab, N, and E genes of the SARS-CoV-2 genome are targeted in the RT-PCR assay (Figure 1A). Primers and TaqMan probes designed for conserved regions of the SARS-CoV-2 virus genome allow specific amplification and detection of the viral RNA from all strains of SARS-CoV-2 from respiratory specimens. The Human RNase P gene is used as an Internal Control (IC) to monitor viral RNA extraction efficiency and assess amplifiable RNA in the samples to be tested. The test is a multiplex RT-PCR assay consisting of one reaction with primers and probes for the viral gene targets (Orf1ab, N and E genes) and IC in one tube, designed to increase assay throughput.

Figure 1.
  • Download figure
  • Open in new tab
Figure 1.

(A) SARS-CoV-2 genome structure and assay target genes. (B) a high throughput workflow for SARS-COV-2 detection from sample collection to result availability within about 4 hrs.

We demonstrate here that saliva sampling is an adequate alternative to NPS and OPS sampling and can be used for COVID-19 testing using the QuantiVirus SARS-CoV-2 multiplex test.

Methods

Study design and ethics

Besides contrived saliva samples, deidentified leftover patient NPS and saliva samples were used in the study. All patient specimens were collected in June-September 2020 and previously tested at UCSF affiliated San Francisco VAMC clinical laboratories and DiaCarta clinical laboratory for clinical diagnostic or screening purpose. Other than qualitative RT-PCR results (positive or negative), only PCR cycle threshold (Ct) values were included in study analysis and no patient clinical chart reviews were performed. This study was approved by the institutional review board (IRB) at UCSF (UCSF IRB #11-05207) as a no-subject contact study with waiver of consent and as exempt under category 4.

Clinical specimens

Clinical samples were collected from patients who had previously been tested positive for SARS-CoV-2. Paired NPS and saliva samples were collected at the same time. The QuantiVirus™Saliva Collection Kit (DiaCarta, Inc. cat# DC-11-0021) was used for saliva collection, following the kit insert instructions and under the supervision of healthcare providers. Each saliva sample contains 2 mL liquid saliva and 2 mL viral transport media. The NPS and saliva samples are refrigerated and processed for testing within 24 hours after collection.

Sample pooling

Patient saliva and healthy saliva samples were pooled together according to the experiment design for 1:5 (i.e., 1 positive mixed with 5 negatives) and 1:11 (i.e., 1 positive mixed with 11 negatives) pooling. After mixing the samples, the viral RNA was extracted according to the following protocol.

Viral RNA extraction

MGI’s automatic RNA/DNA extraction instrument MGISP-960 or Thermo PureLink™ Viral RNA/DNA Mini Kit (Cat. 12280050) was used for the SARS-CoV-2 viral RNA extraction according to the manufacturer’s instructions, for which 200 µL of each NPS or saliva sample was used. For each batch of clinical samples to be tested, an extraction control (EC) was included (spike 20 µL of EC from the QuantiVirus™ SARS-CoV-2 multiplex kit into 180 µL sterile RNase-free water). The clinical samples and spiked EC were processed and extracted on the MGI platform. The extraction output is RNA in 30-50 µL RNase-free water, 5.5 µL of which is used for the PCR reaction per test. The turnaround time from sample extraction to PCR final report is around 4 hrs (Figure 1B). Precautions were taken while handling extracted RNA samples to avoid RNA degradation. Extracted RNA samples were stored at -80°C if not immediately used for RT-PCR.

Multiplex primer and probe design

Target gene sequences in the SARS-CoV-2 genome, the N gene, E gene and ORF1ab gene were identified and selected for test development. The gene sequences were retrieved from GenBank and GISAID databases for primer and probe designs to ensure coverage of all SARS-CoV-2 strains. Multiple alignments of the collected sequences were performed using CLC Main Workbench 20.0.4., and conserved regions in each target gene were identified using BioEditor 7.2.5. prior to primer and probe designs. Primers and probes were designed to target the most conserved regions of each of the target genes of the viral genome, using Primer3plus software and following general rules of real-time PCR design. All primers were designed with a melting temperature (Tm) of approximately 60°C and the probes were designed with a Tm of about 65°C. The amplicon sizes were kept as short as possible within the range of 70 bp to 150 bp for each primer pair to achieve better amplification efficiency and detection sensitivity. All primers and probes were synthesized by Integrated DNA Technologies, Inc. (IDT, Coralville, IA, USA) and LGC Biosearch Technologies (Novato, CA, USA), respectively.

Real-time reverse-transcription PCR (rRT-PCR)

The total volume of one RT-PCR reaction for all targets is 10 µL, including 5.5 µL of RNA, 2.0 μL of 5x primer and probe mixture (final concentration of 0.2 µM and 0.1 µM, respectively), and 2.5 µL of 4x TaqPath™ 1-Step RT-qPCR Master Mix (Catalog number A28526, Thermo Fisher, Waltham, MA) or 4x Inhibitor-Tolerant RT-qPCR mix (MDX016-50, Meridian Bioscience, Tennessee). Thermal cycling was performed at 25°C for 2 min for UNG incubation and 53°C for 10 min for reverse transcription, followed by 95°C for 2 min and then 45 cycles of 95°C for 3 sec, and 60°C for 30 sec. QuantStudio™ 5 Real-Time PCR System (Thermo Fisher, USA), Applied Biosystems™ 7500 Fast Dx Real-Time PCR Instrument (Thermo Fisher, USA), BioRad CFX384 (Bio-Rad, USA) and Roche LightCycler 480 II (Roche, USA) were used for rRT-PCR amplification and detection.

Microorganism Panel for cross-reactivity

MERS-coronavirus, SARS-CoV coronavirus samples were ordered from IDT. NATtrol Respiratory Validation Panel was ordered from ZeptoMetrix (cat# NATRVP-3, Buffalo, NY). RNA/DNA were extracted from high titer stocks of the potentially cross-reacting microorganisms.

Statistical data analysis

Average cycle threshold (Ct), standard deviation (SD) and coefficient of variation (CV) were calculated using Microsoft Office Excel 365 software (Microsoft, Redmond, WA). Clinical sensitivity, clinical specificity, Positive Predictive Value (PPV) and Negative Predictive value (NPV) at two-sided 95% confidence interval (CI) were analyzed using MedCalc software Version 19.3.1. ROC curve was plotted by R package pROC. NP and saliva pair analysis was conducted by Wilcoxon signed rank test.

Results

Validation of QuantiVirus ™SARS-CoV-2 Multiplex Test kit

Analytical sensitivity

We validated the QuantiVirus™SARS-CoV-2 Multiplex kit on four qPCR instruments from different vendors, using contrived saliva samples. The overall analytical sensitivity (lower limit of detection or LOD) is around 100-200 copies/mL.

The validation data established that the LOD of the assay is 200 copies/mL on ABI 7500 Fast Dx (Table 1a), 100 copies/mL on Bio-Rad CFX384 (Table 1b), 200 copies/mL on Roche LightCycler 480 II (Table 1c), and 200 copies/mL on the Thermo Fisher QuantStudio 5 (Table 1d).

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1a.

Summary of twenty replicates for analytical sensitivity confirmation on the ABI 7500 Dx

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1b.

Summary of twenty replicates for analytical sensitivity confirmation on the BioRad CFX 384

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1c.

Summary of twenty replicates for analytical sensitivity confirmation on the Roche LC 480

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 1d.

Summary of twenty replicates for analytical sensitivity confirmation on the ABI QS5

Cross-reactivity (assay specificity)

We tested the cross-reactivity as part of the assay development. MERS-coronavirus and SARS-CoV coronavirus samples were ordered from IDT and NATtrol Respiratory Verification Panel from ZeptoMetrix (cat#NATRVP-3). RNA/DNA were extracted from high titer stocks of the potentially cross-reacting microorganisms (estimated 1.0E+05 units/mL), RNA/DNA were extracted from 100 µL microorganisms’ stocks using the Thermo Fisher viral RNA extraction kit (PureLink™ Viral RNA/DNA Mini Kit) or Qiagen QIAamp DNA Mini Kit. Extracted RNA/DNA were eluted to 100 μL with sterile RNase-free water. 5.5 μL of the purified RNA/DNA samples was used for each reaction and tested using the QuantiVirus ™SARS-CoV-2 multiplex Test Kit. The cross-reactivity testing results are summarized in Table 2.

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 2.

Summary of cross-reactivity evaluation of the QuantiVirus ™SARS CoV-2 Test

The cross-reactivity tests were run in triplicate, and all test controls passed (Positive controls Ct<25, No target control Ct >45, Extraction control has RP Ct ∼28). The tested organisms were all negative for the targeted N, E and ORF1ab genes of SARS-CoV-2, indicating there is no cross-reactivity between SARS-CoV-2 primers & probes and any of the comparison organisms tested. The cross reactivity with common Human coronaviruses and MERS-coronavirus was also tested, and there was no cross reactivity at 105 PFU/mL.

Assay sensitivity verification

We spiked non-infectious viral particles (SeraCare AccuPlex SARS-CoV-2 Reference Material Kit, Cat # 0505-0126) into healthy donor saliva which were confirmed by SARS-CoV-2 qPCR test to be negative, and tested each using three different qPCR instruments, ABI QuantStudio 5, ABI 7500 Fast Dx and BioRad CFX 384.

Clinical evaluation of the QuantiVirus™ SARS-CoV-2 Multiplex Test was conducted with saliva specimens including 40 contrived positive and 30 negative samples. Saliva samples collected from healthy donors were mixed with the lysis buffer at 1:1 ratio before spiking in non-infectious viral particles (SeraCare AccuPlex SARS-CoV-2 Reference Material Kit) (Table 3).

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 3a.

Evaluation of contrived clinical sample with viral particles (Bio-Rad CFX 384)

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 3b.

Evaluation of contrived clinical sample with viral particles (ABI QuantStudio 5)

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 3c.

Evaluation of contrived clinical sample with viral particles (ABI 7500 Fast Dx)

20 contrived positive saliva samples were created with the addition of non-infectious viral particles templates at 1x estimated LOD (1×200 copies/mL), 10 saliva samples were spiked at 1.5xLoD (300 copies/mL), and another 10 saliva samples were spiked at 2.5xLoD (500 copies/mL). Viral RNA was extracted from spiked samples and tested with the QuantiVirus™ SARS-CoV-2 multiplex test kit. The results showed that all 40 spiked saliva samples tested positive and all 30 control saliva samples tested negative on all three PCR instruments (Tables 3a, b and c) with PPA being 100% (95% CI: 0.76-0.99) and NPA of 100% (95% CI: 0.91-1.00).

Clinical evaluation on NPS samples

Using the QuantiVirus™ SARS-CoV-2 Multiplex Test, we tested clinical NPS samples including 85 positive samples and 90 negative samples which previously had been tested on Abbott m2000 molecular system using Abbott Real-Time SARS-CoV-2 testing kits (Table 4). The data shows that the clinical sensitivity of QuantiVirus™ SARS-CoV-2 multiplex test is 98.8% (95% CI: 92.7% -99.9%) and specificity is 100% (95% CI: 94.9%-100%). Its PPV is 100% (95% CI: 94.6%-100%) and NPV is 98.9% (95% CI: 93.1%-99.9%). ROC curve shows its AUC = ∼0.988 for this sensitivity and specificity (Figure 2).

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 4.

Clinical Sample Evaluation with QuantiVirus™ SARS-CoV-2 Multiplex Test

Figure 2.
  • Download figure
  • Open in new tab
Figure 2.

ROC curve of the QuantiVirus ™SARS-CoV-2 multiplex test for detecting SARS-CoV-2 using NPS samples, with specificity data on X axis and sensitivity on Y axis. AUC (Area Under Curve) was calculated as 0.988. This figure was plotted by R package pROC.

Clinical evaluation of paired NPS and saliva samples

We tested and evaluated the concordance between paired NPS and saliva samples collected from patients by QuantiVirus™ SARS-CoV-2 multiplex tests. Among the 20 pairs of nasopharyngeal swabs (NP) and saliva samples, the test results were the same for 16 pairs (16/20, 80% concordance rate), including 5 positive pairs and 11 negative pairs. There were four samples for which the results were discordant (Table 5). Of these four pairs, one pair was NPS positive and saliva negative, whereas the other three pairs were NPS negative and saliva positive. Nevertheless, we compared NPS and saliva specimens by Wilcoxon matched pair signed rank test. The two samples types show 80% concordance with no significant differences (p=0.13, Figure 3a), and its RP were similar between two types of specimens (p=0.06, Figure 3b).

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 5.

Paired NPS and Saliva Tested by QuantiVirus ™SARS-CoV-2 Multiplex Test

Figure 3.
  • Download figure
  • Open in new tab
Figure 3.

Clinical evaluation of paired nasopharyngeal and saliva samples. Cycle threshold (Ct) values for target gene and RP gene of NPS and saliva specimens were compared by Wilcoxon matched pairs signed rank test. Cycle threshold (Ct) values for viral E gene, N gene, O gene (ORF1ab), and human RNase P gene (RP) for NP and saliva specimens. A) E, N, and O Ct values for paired NP and saliva samples. Pairs are connected by a line. The Ct was set to 42 for samples in which signal was not detected. Ct values of E, N, and O were comparable between the two types of samples by Wilcoxon signed rank test. NPS and saliva concordance is about 80% with no significant differences (p=0.13). B) RP Ct values for NP and saliva specimens were similar between the two types of samples by Wilcoxon signed rank test.

Comparison of QuantiVirus™ SARS-CoV-2 multiplex kit with FDA EUA approved Abbott Realtime SARS-CoV-2 kit

We tested 24 saliva samples of recovering COVID-19 patients with the QuantiVirus™ SARS-CoV-2 multiplex kit in comparison with the Abbott m2000 SARS-CoV-2 PCR kit in parallel (Table 6). Data showed a concordance of the assays of about 88%. There were three samples detected by QuantiVirus™ SARS-CoV-2 multiplex kit, but not detectable with the Abbott kit (patients #8, 11 and 12), consistent with the reported higher sensitivity of QuantiVirus™ PCR assay.22

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 6.

Comparison of Abbott m2000 SARS-CoV-2 PCR test and DiaCarta QuantiVirus™SARS-CoV-2 multiplex PCR test

Population screening using saliva samples

We tested 389 total saliva specimens from the general population of asymptomatic people in Los Angeles and the San Francisco Bay Area. The screened population was represented by African Americans, White, Asian, and Latinx, with ages ranging from 18 to 80 (average 41) years old. From May 8 to Aug 26, 2020, 301 saliva samples were tested, and 5 samples were tested positive for SARS-CoV-2 by the QuantiVirus ™SARS-CoV-2 multiplex PCR test. The 5 positives corresponded to 4 males of ages 19, 51, 52 and 54, and 1 female of age 34. Overall detection rate was 1.66% (Table 7). In another recent testing run of 88 saliva samples, 2 samples were positive and 86 were negative, with an overall positive detection rate of 2.27%. Overall, we had screened 389 people from the general population and found that 7 people were positive for SARS-CoV-2 with a detection rate of 1.8%, consistent with the reported average positive test rate from the same periods in the two metropolitan regions.

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 7.

Summary of saliva-based COVID-19 screening using QuantiVirus ™SARS-CoV-2 multiplex PCR test

Evaluation of pooling saliva samples for SARS-CoV-2 screening

To test the feasibility of pooling saliva specimens for screening asymptomatic patients, we pooled healthy donor saliva (negative) and patient saliva (positive) and tested a total of 77 pooled samples (1 patient sample mixed with 5 healthy saliva samples) and 54 pooled health samples (mixed 6 health samples) (Table 8). Of the 77 pooled saliva samples, 73 were tested positive (average Ct of three genes: O gene Ct∼29.8; E gene 30.9 and N gene Ct ∼31,0) and 4 was negative. The average IC RP Ct was 21.9 for all 131 pooled samples. Positive Predictive Value (PPV) is 100% (95% CI: 93.8%-100%). Negative Predictive Value (NPV) is 93.1% (95% CI: 82.5-97.8%). Additionally, we tested a total of 49 pooled saliva samples, created by mixing 1 patient sample with 11 healthy samples. Of the 49 pooled samples, 44 were detected positive (O gene, E gene and N gene average Ct 31.8, 32.1 and 31.9) and 5 was negative. Its IC RP average Ct was 22.3 for all 49 pooled saliva samples and additional 20 pooled healthy saliva samples. PPV is 100% (95% CI: 89.9%-100%) and NPV is 80.0% (95% CI: 58.7%-92.4%), respectively.

View this table:
  • View inline
  • View popup
  • Download powerpoint
Table 8.

Saliva sample pooling for SARS-CoV-2 detection by QuantiVirus ™SARS-COV-2 multiplex PCR test

Discussion

We have developed and validated a multiplex rRT-PCR assay for SARS-CoV-2 detection with clinical sensitivity being 98.8% (95% CI: 92.7%-99.9%) and specificity of 100% (95% CI: 94.9%-100%). Its PPV is 100% (95% CI: 94.6%-100%) and NPV is 98.9% (95% CI: 93.1%-99.9%). The ROC curve shows an AUC of ∼0.988. The detection of three viral target genes in one PCR tube enables a high throughput test using RT-qPCR. For these validated PCR platforms, 381 patient samples can be tested in each run (plus 3 controls). We have developed and integrated MGISP-960 High-throughput Automated Sample Preparation System which can extract 192 samples (2×96) in about 80 min. For CLIA labs with two MGI-960 machines, 380 samples can be tested with results available within 4 hrs. (Figure 1b)

We spiked SARS-CoV-2 viral particles into healthy donor saliva and confirmed that the analytical sensitivity (LOD) of the QuantiVirus™ multiplex RT-qPCR test is ∼100 copies/mL for Bio-Rad CFX 384 and ∼200 copies/mL for ABI QS5, ABI 7500Dx and Roche LC 480. The multiplex RT-qPCR test can simultaneously detect three viral gene targets, which can minimize false negative results as chances of simultaneous mutations in all three target genes in the viral genome are highly unlikely. On the other hand, this also confirms that human saliva samples do not inhibit the RT-qPCR reaction possibly due to the fact that inhibitor-tolerant RT-PCR master mix was used in the QuantiVirus™ SARS-CoV-2 test kit.

Leung et al 6 analyzed 95 patient-matched paired samples from 62 patients including 29 confirmed patients with COVID-19 and 33 COVID-19 negative patients. The concordance rate was 78.9% (75/95 samples) between NP and saliva. Vogels et al 19 reported a positive agreement of 83.8% (31/37 positive samples) for nasopharyngeal swabs and saliva when using TaqPath COVID-19 combo kit. Our data showed 80% concordance and no significant differences between NP and saliva, which is consistent with Leung’s & Vogels’s reports. Interestingly, for patient #4, the viral RNA was detected in the NPS sample by both Abbott test and DiaCarta test, but viral RNA was not detected in the saliva by either test. For patient #6, viral RNA was not detected in the NPS sample by either test, whereas viral RNA was detected in the saliva by both tests. This observation suggests that sample collection variables such as the time of collection and NPS sample quality do matter in SARS-CoV-2 testing.

The 20 paired NPS-saliva samples were collected from recovering patients being evaluated prior to their release from self-quarantine. Consequently, their viral loads were much lower (100-1000 copies/mL), compared to the viral loads expected for the initial diagnostic testing. For patients with active SARS-CoV-2 infection (viral loads typically are above 10,000 copies/mL), there should be no problem detecting the virus in adequately collected saliva samples.

Landry et al 20 described that most of saliva samples from sick patients were thick, stringy, and difficult to pipet. Since we used the QuantiVirus™ SARS-CoV-2 saliva sample collection kit which has VTM solution in the collection tube, the saliva was diluted 50% and therefore much easier to process. Matic et al 21 used PBS at a 1:2 dilution that also helped resolve highly viscous saliva samples; however, manual dilution after collection may be associated with pipetting errors and cross contamination.

The QuantiVirus™ SARS-CoV-2 multiplex test results were 87.5% in concordance with FDA EUA approved Abbott RealTime SARS-CoV-2 results for saliva samples, with a higher detection rate overall. In fact, this observation is consistent with recently reported test sensitivity among various SARS-CoV-2 molecular tests. FDA published its SARS-CoV-2 Reference Panel Comparative Data on its website on Sept 15th, 2020 22. It reported that QuantiVirus™ SARS-CoV-2 multiplex Kit has LOD of 600 NDU/mL whereas Abbott Realtime SARS-CoV-2 assay has LOD of 5400 NDU/mL. Accordingly, the reason for the observation that SARS-CoV-2 viral RNA was detected in three patient samples by the QuantiVirus™ SARS-CoV-2 multiplex test but not by Abbott Realtime SARS-CoV-2 assay was likely due to the higher sensitivity of the QuantiVirus™ SARS-CoV-2 multiplex assay. It also demonstrated that saliva specimens represent a viable specimen type that can be easily applied for COVID-19 testing when using more sensitive tests.

A total of 389 saliva specimens from the general population were tested and demonstrated the feasibility of using saliva for large scale population screening. Saliva is a non-invasive and easily collectable specimen for COVID-19 screening. Given the drawbacks of nasopharyngeal and oropharyngeal swab sample collection, saliva sampling could be applied as an acceptable alternative 23.

With saliva pooling strategy, we have demonstrated that 6-samples pooling (1 patient mixed with 5 healthy saliva samples) has 94.8% sensitivity (95% CI: 86.5-98.3%) and 100% specificity (95% CI:91.7-100%), As noted, of the 77 pooled saliva samples, 4 pooling samples were tested negative. In fact, for these 4 pooled samples, the individual patient samples used for the pooling had Ct of 34.4, 34.8, 35.7 and 37.5 for ORF1ab gene, respectively, consistent with low viral loads (less than 100-200 copies/mL) (see Table 1a-d). Therefore, in order to detect weakly positive patient in pooled samples, a RT-PCR test with LOD at 100-200 copies/mL or better is required. If pooling testing is considered, each clinical laboratory should establish laboratory-specific pooling protocol based on the LOD of SARS-CoV-2 molecular test.

In summary, we have demonstrated that saliva specimens can be reliably used for SARS-CoV-2 detection, and saliva-based large-scale population screening for COVID-19 with or without pooling is feasible.

Data Availability

the data are available for public

Acknowledgements

The authors thank Dr. Ramanathan Vairavan and Eric Chu for reviewing and editing this manuscript, and Christopher Teixeira, Fady Ettnas, Enas Eltarazy and Eric Abbott for their technical contributions.

References

  1. 1.↵
    Zhu N, Zhang D, Wang W, Li X, Yang B, Song J, et al. A Novel Coronavirus from Patients with Pneumonia in China, 2019. N Engl J Med 382 (2020) 727–733
    OpenUrlCrossRefPubMed
  2. 2.↵
    World Health Organization (WHO). Coronavirus. Geneva: WHO; 2020 [Accessed 21 Jan 2020]. Available from: https://www.who.int/health-topics/coronavirus
  3. 3.↵
    Li Q, Guan X, Wu P, Wang X, Zhou L, Tong Y, et al. Early Transmission Dynamics in Wuhan, China, of Novel Coronavirus-Infected Pneumonia. N Engl J Med. 382 (2020) 1199–1207
    OpenUrlCrossRefPubMed
  4. 4.
    Chen N, Zhou M, Dong X, Qu J, Gong F, Han Y, et al. Epidemiological and clinical characteristics of 99 cases of 2019 novel coronavirus pneumonia in Wuhan, China: a descriptive study. Lancet. 395(2020) 507–514.
    OpenUrlCrossRefPubMed
  5. 5.↵
    Chan JF, Yuan S, Kok KH, To KK, Chu H, Yang J, et al. A familial cluster of pneumonia associated with the 2019 novel coronavirus indicating person-to-person transmission: a study of a family cluster. Lancet. 395 (2020) 514–523.
    OpenUrlCrossRefPubMed
  6. 6.↵
    Leung EC, Chow VC, Lee MK and Lai RW. Deep throat saliva as an alternative diagnostic specimen type for the detection of SARS-CoV-2. J Med Virol. (2020)1–4 https://doi.org/10.1002/jmv.26258
  7. 7.↵
    Wölfel R, Corman VM, Guggemos W, Seilmaier M, Zange S, Müller MA et al. Virological assessment of hospitalized patients with COVID-2019. Nature 581 (2020) 465–469
    OpenUrlCrossRefPubMed
  8. 8.
    He X, Lau EH, Wu P, Deng X, Wang J, Hao X et al. Temporal dynamics in viral shedding and transmissibility of COVID-19. Nat Med 26 (2020) 672–675.
    OpenUrlPubMed
  9. 9.
    Nagura-Ikeda M, Imai K, Tabata S, Miyoshi K, Murahara N, Mizuno T et al. Clinical evaluation of self-collected saliva by RT-qPCR, direct RT-qPCR, RT-LAMP, and a rapid antigen test to diagnose COVID-19. J Clin Microbiol. 58 (2020) e01438–20
    OpenUrl
  10. 10.↵
    Kinloch N, Ritchie G, Brumme CJ, Dong W, Dong W, Lawson T et al. Suboptimal biological sampling as a probable cause of false-negative COVID-19 diagnostic test results. J Infect Dis 222(2020) 899–902
    OpenUrl
  11. 11.↵
    Wyllie A, Fournier J, Casanovas-Massana A, Campbell M, Tokuyama M, Vijayakumar P. et al. Saliva is more sensitive for SARS-CoV-2 detection in COVID-19 patients than nasopharyngeal swabs. medRxiv (2020) doi: https://doi.org/10.1101/2020.04.16.20067835
  12. 12.↵
    Hanson KE, Barker AP, Hillyard D.R, Gilmore N, Barrett JW, Orlandi RR et al. Self-collected anterior nasal and saliva specimens versus healthcare worker-collected nasopharyngeal swabs for the molecular detection of SARS-CoV-2. J. Clin. Microbiol. (2020). Doi:10.1128/JCM.012824-20
  13. 13.↵
    Centre for Health Protection. Severe respiratory disease associated with a novel infectious agent—letters to doctors. (2020). Accessed: February 5, 2020: https://www.chp.gov.hk/en/miniweb/letters/100063.html.
  14. 14.↵
    FDA EUA, Individual EUAs for Molecular Diagnostic Tests for SARS-CoV-2 https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/vitro-diagnostics-euas#umbrella-molecular
  15. 15.↵
    Sri Santosh T, Parmar R, Anand H, Srikanth K, Saritha M. A Review of Salivary Diagnostics and Its Potential Implication in Detection of Covid-19. Cureus 12 (2020) e7708. DOI 10.7759/cureus.7708
    OpenUrlCrossRef
  16. 16.
    To KK, Tsang OT, Yip CC, Chan KH, Wu TC, Chan JM. et al. Consistent Detection of 2019 Novel Coronavirus in Saliva. Clin. Infect. Dis. 71 (2020) 841–843. doi:10.1093/cid/ciaa149.
    OpenUrlCrossRefPubMed
  17. 17.
    Chen L, Zhao J, Peng J, Li X, Deng X, Geng Z. et al. Detection of 2019-nCoV in Saliva and Characterization of Oral Symptoms in COVID-19 Patients. (2020) https://ssrn.com/abstract=3557140
  18. 18.↵
    Xu R, Cui B, Duan X, Zhang P, Zhou X and Yuan Q. 2020. Saliva: potential diagnostic value and transmission of 2019-nCoV. International Journal of Oral Science 12 (2020):11 https://doi.org/10.1038/s41368-020-0080-z
    OpenUrl
  19. 19.↵
    Vogels C, Watkins A, Harden C, Brackney D, Shafer J, Wang J. et al. SalivaDirect: Simple and sensitive molecular diagnostic test for SARS-COV-2 surveillance. MedRxiv (2020) https://doi.org/10.1101/2020.08.03.20167791
  20. 20.↵
    Landy M, Criscuolo J and Peaper D. Challenges in use of saliva for detection of SARS-COV-2 RNA in symptomatic outpatients. J. Clincal Virology 130 (2020) 104567.
    OpenUrl
  21. 21.↵
    Matic N, Lawson T, Ritchie G, Stefanovic A, Leung V, Champagne S. et al./person-group> Automated molecular testing of saliva for SARS-COV-2 detection. MedRxiV (2020) https://doi.org/10.1101/2020.08.11.20170613
  22. 22.↵
    FDA SARS-CoV-2 Reference Panel Comparative Data https://www.fda.gov/medical-devices/coronavirus-covid-19-and-medical-devices/sars-cov-2-reference-panel-comparative-data?from=singlemessage&isappinstalled=0
  23. 23.↵
    Fakheran O, Dehghannejad M and Khademi A. Saliva as a diagnostic specimen for detection of SARS-CoV-2 in suspected patients: a scoping review. Infectious Diseases of Poverty 9 (2020) 100–107 https://doi.org/10.1186/s40249-020-00728-w
    OpenUrl
Back to top
PreviousNext
Posted November 01, 2020.
Download PDF
Data/Code
Email

Thank you for your interest in spreading the word about medRxiv.

NOTE: Your email address is requested solely to identify you as the sender of this article.

Enter multiple addresses on separate lines or separate them with commas.
Saliva as a testing specimen with or without pooling for SARS-CoV-2 detection by multiplex RT-PCR test
(Your Name) has forwarded a page to you from medRxiv
(Your Name) thought you would like to see this page from the medRxiv website.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Share
Saliva as a testing specimen with or without pooling for SARS-CoV-2 detection by multiplex RT-PCR test
Qing Sun, Jonathan Li, Hui Ren, Larry Pastor, Yulia Loginova, Roberta Madej, Kristopher Taylor, Joseph K. Wong, Zhao Zhang, Aiguo Zhang, Chuanyi M. Lu, Michael Y. Sha
medRxiv 2020.10.27.20219196; doi: https://doi.org/10.1101/2020.10.27.20219196
Twitter logo Facebook logo LinkedIn logo Mendeley logo
Citation Tools
Saliva as a testing specimen with or without pooling for SARS-CoV-2 detection by multiplex RT-PCR test
Qing Sun, Jonathan Li, Hui Ren, Larry Pastor, Yulia Loginova, Roberta Madej, Kristopher Taylor, Joseph K. Wong, Zhao Zhang, Aiguo Zhang, Chuanyi M. Lu, Michael Y. Sha
medRxiv 2020.10.27.20219196; doi: https://doi.org/10.1101/2020.10.27.20219196

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Subject Area

  • Infectious Diseases (except HIV/AIDS)
Subject Areas
All Articles
  • Addiction Medicine (349)
  • Allergy and Immunology (668)
  • Allergy and Immunology (668)
  • Anesthesia (181)
  • Cardiovascular Medicine (2648)
  • Dentistry and Oral Medicine (316)
  • Dermatology (223)
  • Emergency Medicine (399)
  • Endocrinology (including Diabetes Mellitus and Metabolic Disease) (942)
  • Epidemiology (12228)
  • Forensic Medicine (10)
  • Gastroenterology (759)
  • Genetic and Genomic Medicine (4103)
  • Geriatric Medicine (387)
  • Health Economics (680)
  • Health Informatics (2657)
  • Health Policy (1005)
  • Health Systems and Quality Improvement (985)
  • Hematology (363)
  • HIV/AIDS (851)
  • Infectious Diseases (except HIV/AIDS) (13695)
  • Intensive Care and Critical Care Medicine (797)
  • Medical Education (399)
  • Medical Ethics (109)
  • Nephrology (436)
  • Neurology (3882)
  • Nursing (209)
  • Nutrition (577)
  • Obstetrics and Gynecology (739)
  • Occupational and Environmental Health (695)
  • Oncology (2030)
  • Ophthalmology (585)
  • Orthopedics (240)
  • Otolaryngology (306)
  • Pain Medicine (250)
  • Palliative Medicine (75)
  • Pathology (473)
  • Pediatrics (1115)
  • Pharmacology and Therapeutics (466)
  • Primary Care Research (452)
  • Psychiatry and Clinical Psychology (3432)
  • Public and Global Health (6527)
  • Radiology and Imaging (1403)
  • Rehabilitation Medicine and Physical Therapy (814)
  • Respiratory Medicine (871)
  • Rheumatology (409)
  • Sexual and Reproductive Health (410)
  • Sports Medicine (342)
  • Surgery (448)
  • Toxicology (53)
  • Transplantation (185)
  • Urology (165)