Abstract
Introduction The handheld dynamometer has been validated to measure muscle strength in different muscle groups. However, to date, it has not been tested in individuals who experience pain induced by hip osteoarthritis. The current study aimed to evaluate the intra- and inter-rater reliability, agreement, and minimal detectable change of the Lafayette handheld dynamometer, model 1165, to assess the peak force (Pk) and average peak force (Af) of hip muscles in individuals with symptomatic hip osteoarthritis.
Methods Twenty participants with hip osteoarthritis (mean ± SD age: 58.7±15.3 years; body mass index: 28.8±4.2 kg/m2) and a pain intensity on the Visual Analogue Scale0020 ≥ 4 (8.05±1.2) were recruited to participate in this study. Pk and Af of hip flexors (seated position), abductors and adductors (supine position), and extensors (prone position) were collected in a single day by two independent raters, each one obtaining test and retest in randomly ordered separate sessions.
Results The intra-rater intraclass correlation coefficient (ICC) was classified as good (>0.75) or excellent (≥0.90) for all muscle groups, and all inter-rater ICCs were classified as excellent. Rater A had a lower standard error of measurement compared to rater B, ranging from 0.15 to 0.58 kilogram-force (Kgf) compared with 0.34 to 1.25 kg, respectively. However, the inter-rater comparison showed a minimal detectable change < 10% for all Pk and Af measures (except Af for the abductor muscle group). Finally, the inter-rater Bland-Altman analysis demonstrated good agreement for abductors, adductors, and extensors.
Conclusion Despite pain and dysfunction related to hip osteoarthritis, the handheld dynamometer was shown to be a reliable tool to assess hip muscle strength, with good to excellent intra- and inter-rater ICCs, satisfactory agreement, and small values for minimal detectable change.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
The author(s) received no specific funding for this work.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
This study was approved by the local Ethics Committee (CAAE 40347320.1.1001.0025).
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data may be acessed at https://drive.google.com/drive/folders/1nxnIzbPSkf-UBJSeAloT3WQMvO-GRO2j?usp=sharing.
https://drive.google.com/drive/folders/1nxnIzbPSkf-UBJSeAloT3WQMvO-GRO2j?usp=sharing