Skip to main content
medRxiv
  • Home
  • About
  • Submit
  • ALERTS / RSS
Advanced Search

Initial Experience with Intercostal Insertion of an Extravascular ICD Lead Compatible with Existing Pulse Generators

Martin C. Burke, View ORCID ProfileReinoud E. Knops, View ORCID ProfileVivek Reddy, Johan Aasbo, View ORCID ProfileMichael Husby, View ORCID ProfileAlan Marcovecchio, Mark O’Connor, View ORCID ProfileRick Sanghera, Don Scheck, Shari Pepplinkhuizen, Adrian Ebner
doi: https://doi.org/10.1101/2023.02.21.23286264
Martin C. Burke
aCorVita Science Foundation, Chicago, Illinois
bDepartment of Clinical and Experimental Cardiology, Amsterdam UMC, Academic Medical Center, Amsterdam, the Netherlands
eAtaCor Medical, Inc., San Clemente, California
DO
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: mburke{at}corvitahealth.org
Reinoud E. Knops
bDepartment of Clinical and Experimental Cardiology, Amsterdam UMC, Academic Medical Center, Amsterdam, the Netherlands
MD, PhD
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Reinoud E. Knops
Vivek Reddy
cMount Sinai Medical School, New York, NY
MD
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Vivek Reddy
Johan Aasbo
dLexington Cardiology and Baptist Health, Lexington, KY
DO
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Michael Husby
eAtaCor Medical, Inc., San Clemente, California
MS, MPH
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Michael Husby
Alan Marcovecchio
eAtaCor Medical, Inc., San Clemente, California
MSc
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Alan Marcovecchio
Mark O’Connor
eAtaCor Medical, Inc., San Clemente, California
BS
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Rick Sanghera
eAtaCor Medical, Inc., San Clemente, California
BSc, MBA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Rick Sanghera
Don Scheck
eAtaCor Medical, Inc., San Clemente, California
BS
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Shari Pepplinkhuizen
bDepartment of Clinical and Experimental Cardiology, Amsterdam UMC, Academic Medical Center, Amsterdam, the Netherlands
MD
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Adrian Ebner
fCardiovascular Department, Sanatorio Italiano, Asunción, Paraguay
MD
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Abstract
  • Full Text
  • Info/History
  • Metrics
  • Data/Code
  • Preview PDF
Loading

Structured Abstract

Objectives This study assessed safety and feasibility of a novel extravascular (EV) implantable cardioverter-defibrillator (ICD) lead when inserted anteriorly through a rib space and connected to various commercially available ICD pulse generators (PGs) placed in either a left mid-axillary or left pectoral pocket.

Background Currently available or investigational, EV-ICDs include a subcutaneous or subxiphoid lead connected to customized EV-ICD PGs. This novel EV-ICD (AtaCor Medical, Inc, San Clemente, CA) employs a unique intercostal implant technique and is designed to function with commercial DF-4 ICD PGs.

Methods In this non-randomized, single-center, acute study, 36 de novo or replacement transvenous (TV) ICD patients enrolled to receive a concomitant EV-ICD lead inserted through an intercostal space along the left parasternal margin. EV-ICD leads were connected to DF-4 compatible ICD PGs positioned in either a left mid-axillary or pectoral pocket for acute sensing and defibrillation testing. Defibrillation testing started at 30 Joules (J) and stepped down in 10 J increments following conversion success and stepped up in 5 J increments following conversion failure.

Results Successful acute defibrillation using ≤ 35 J was noted in 100% of left mid-axillary PG subjects (n=27, mean 16.3 ± 8.6 J) and 83% of left pectoral PG subjects (n=6, mean 21.0 ± 8.4 J). All evaluable episodes (n=93) were automatically sensed, detected, and shocked. No serious device-related intraoperative adverse events were observed.

Conclusions This first-in-human study documented safe and reliable placement of a novel extravascular ICD lead with effective sensing and defibrillation of induced ventricular fibrillation using commercial DF-4 ICD pulse generators.

Condensed Abstract This study assessed feasibility of intercostal implantation of a novel extravascular implantable cardioverter-defibrillator (ICD) lead designed to function with commercial DF-4 pulse generators (PGs). Lead placement was successful in 33 of 36 attempts (94%). Acute defibrillation with ≤35 J was successful in 27 of 27 left mid-axillary PG subjects (100%) and 5 of 6 left pectoral PG subjects (83%). All evaluable episodes (n=93) were automatically detected. No serious device-related intraoperative adverse events were observed. This study demonstrates feasibility of a novel extravascular ICD lead with effective sensing and defibrillating of induced ventricular fibrillation using commercial DF-4 PGs.

Competing Interest Statement

Dr. Burke has received honoraria and research grants from Boston Scientific and AtaCor Medical as well as owns equity in AtaCor Medical. Drs. Knops and Reddy have received honoraria and research grants from Boston Scientific and Medtronic, consult for Abbott, Boston Scientific and AtaCor Medical, and hold equity in AtaCor Medical. Dr. Aasbo have received honoraria and research grants from Boston Scientific and AtaCor Medical. Mr. Husby, Mr. Marcovecchio, Mr. O'Connor, Mr. Sanghera, and Mr. Scheck are employees and shareholders of AtaCor Medical. Dr. Ebner has received research grants from AtaCor Medical.

Clinical Trial

The Parasternal Access for Shocks and Pacing with an Acutely Placed Less-Invasive Lead for EV-ICD (PASS PULL EV ICD) acute study [NCT05099289] and (Phase II) the Sensing and Defibrillation with a Commercially Available ICD Coupled With a Parasternal Extravascular Lead (SECURE EV) acute + 90-day Study [NCT05352776].

Funding Statement

AtaCor Medical is the sponsor of these studies.

Author Declarations

I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.

Yes

The details of the IRB/oversight body that provided approval or exemption for the research described are given below:

Ethics committee Sanatorio Italiano, Paraguay

I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.

Yes

I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).

Yes

I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.

Yes

Footnotes

  • Presented in part as a Late Breaking Clinical Trial at the 2022 Asia-Pacific Heart Rhythm Society Meeting, Singapore, Singapore.

  • Tweet: Early experience inserting a novel, protype ICD lead via the anterior intercostal left parasternal approach into the anterior mediastinum outside the myocardium is feasible anatomically as well as functionally using commercial ICD pulse generators acutely and after 90 days.

  • #Extravascular#ICD#sudden cardiac death.

  • Disclosures: Dr. Burke has received honoraria and research grants from Boston Scientific and AtaCor Medical as well as owns equity in AtaCor Medical. Drs. Knops and Reddy have received honoraria and research grants from Boston Scientific and Medtronic, consult for Abbott, Boston Scientific and AtaCor Medical, and hold equity in AtaCor Medical. Dr. Aasbo have received honoraria and research grants from Boston Scientific and AtaCor Medical. Mr. Husby, Mr. Marcovecchio, Mr. O’Connor, Mr. Sanghera, and Mr. Scheck are employees and shareholders of AtaCor Medical. Dr. Ebner has received research grants from AtaCor Medical.

Data Availability

All data submitted in this manuscript is part of regulatory submissions and will be available online eventually.

  • Abbreviations and Acronyms

    ATP
    Anti-tachycardia Pacing
    EV
    Extravascular
    ICD
    Implantable Cardioverter-Defibrillator
    ICS
    Intercostal Space
    J
    Joules
    PG
    Pulse Generator
    S-ICD
    Subcutaneous ICD
    SCD
    Sudden Cardiac Death
    TV
    Transvenous
  • Copyright 
    The copyright holder for this preprint is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. It is made available under a CC-BY-NC-ND 4.0 International license.
    Back to top
    PreviousNext
    Posted February 26, 2023.
    Download PDF
    Data/Code
    Email

    Thank you for your interest in spreading the word about medRxiv.

    NOTE: Your email address is requested solely to identify you as the sender of this article.

    Enter multiple addresses on separate lines or separate them with commas.
    Initial Experience with Intercostal Insertion of an Extravascular ICD Lead Compatible with Existing Pulse Generators
    (Your Name) has forwarded a page to you from medRxiv
    (Your Name) thought you would like to see this page from the medRxiv website.
    CAPTCHA
    This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
    Share
    Initial Experience with Intercostal Insertion of an Extravascular ICD Lead Compatible with Existing Pulse Generators
    Martin C. Burke, Reinoud E. Knops, Vivek Reddy, Johan Aasbo, Michael Husby, Alan Marcovecchio, Mark O’Connor, Rick Sanghera, Don Scheck, Shari Pepplinkhuizen, Adrian Ebner
    medRxiv 2023.02.21.23286264; doi: https://doi.org/10.1101/2023.02.21.23286264
    Twitter logo Facebook logo LinkedIn logo Mendeley logo
    Citation Tools
    Initial Experience with Intercostal Insertion of an Extravascular ICD Lead Compatible with Existing Pulse Generators
    Martin C. Burke, Reinoud E. Knops, Vivek Reddy, Johan Aasbo, Michael Husby, Alan Marcovecchio, Mark O’Connor, Rick Sanghera, Don Scheck, Shari Pepplinkhuizen, Adrian Ebner
    medRxiv 2023.02.21.23286264; doi: https://doi.org/10.1101/2023.02.21.23286264

    Citation Manager Formats

    • BibTeX
    • Bookends
    • EasyBib
    • EndNote (tagged)
    • EndNote 8 (xml)
    • Medlars
    • Mendeley
    • Papers
    • RefWorks Tagged
    • Ref Manager
    • RIS
    • Zotero
    • Tweet Widget
    • Facebook Like
    • Google Plus One

    Subject Area

    • Cardiovascular Medicine
    Subject Areas
    All Articles
    • Addiction Medicine (349)
    • Allergy and Immunology (668)
    • Allergy and Immunology (668)
    • Anesthesia (181)
    • Cardiovascular Medicine (2648)
    • Dentistry and Oral Medicine (316)
    • Dermatology (223)
    • Emergency Medicine (399)
    • Endocrinology (including Diabetes Mellitus and Metabolic Disease) (942)
    • Epidemiology (12228)
    • Forensic Medicine (10)
    • Gastroenterology (759)
    • Genetic and Genomic Medicine (4103)
    • Geriatric Medicine (387)
    • Health Economics (680)
    • Health Informatics (2657)
    • Health Policy (1005)
    • Health Systems and Quality Improvement (985)
    • Hematology (363)
    • HIV/AIDS (851)
    • Infectious Diseases (except HIV/AIDS) (13695)
    • Intensive Care and Critical Care Medicine (797)
    • Medical Education (399)
    • Medical Ethics (109)
    • Nephrology (436)
    • Neurology (3882)
    • Nursing (209)
    • Nutrition (577)
    • Obstetrics and Gynecology (739)
    • Occupational and Environmental Health (695)
    • Oncology (2030)
    • Ophthalmology (585)
    • Orthopedics (240)
    • Otolaryngology (306)
    • Pain Medicine (250)
    • Palliative Medicine (75)
    • Pathology (473)
    • Pediatrics (1115)
    • Pharmacology and Therapeutics (466)
    • Primary Care Research (452)
    • Psychiatry and Clinical Psychology (3432)
    • Public and Global Health (6527)
    • Radiology and Imaging (1403)
    • Rehabilitation Medicine and Physical Therapy (814)
    • Respiratory Medicine (871)
    • Rheumatology (409)
    • Sexual and Reproductive Health (410)
    • Sports Medicine (342)
    • Surgery (448)
    • Toxicology (53)
    • Transplantation (185)
    • Urology (165)